Next: About this document Up: Use of the TN-7200 Previous: The TN-7200 Comes

Simple Use of the TN-7200

The following steps should be followed for setting up the TN-7200 for use:

  1. Press the SETUP button. Many of the buttons should now be flashing.

  2. Press the ACQUIRE button. A menu of choices will appear on the screen.

  3. Select PHA by pressing ENTER.

    1. Set the conversion gain to 1024.

    2. Set the digital offset to 0.

    3. Set the live time to 20 (this is the counting time in seconds).

    4. Set the preset integral to 0.

  4. Press the SETUP button to stop the setup operation.

  5. Press the button marked 1/4.

  6. Press the ERASE DATA button.

  7. Start an acquisition with the ACQUIRE button.

For further details on using the TN-7200 MCS, please see the Operator's Manual.


wolforth@pet.mni.mcgill.ca
Thu Aug 11 15:18:10 EDT 1994